kic2000炉温测试仪从炉子里出来读不出数据怎么办?
网友回答
【答案】 通常是仪器过热自动保护后怎样进行复位?
退出KIC2000软件
1,找到KIC Profiler 2000\Log\kic2000usersettings.kiccfg ,用记事本方式打开
2,把文件中的ViewHostDialog=0改为=1,然后保存并退出.
再次启动KIC2000软件你就可以看到KICHost界面了.
Can try these steps,remark to using English Language
Remend is use 9V ALKALINE Battery
1.\x09Make sure the SlinKIC2000 software is close.
2.\x09Under the root directory on your hard drive,under the folder the KIC Profiler 2000 you will find a folder called Log.(Using Windows Explorer,open the X:\KIC Profiler 2000\Log\kic2000usersettings.kiccfg file)
3.\x09Under the Log folder there is a file called KIC2000UserSettings.KICCFG
4.\x09Open this file using Note or WordPad text editor.
5.\x09With file opened under the text editor,find the line:( ViewHostDialog=0,Set this line to =1,Save changes to the file.)
6.\x09Restart the SlimKIC2000 software and you motice that a second application will open,the KIC Host.( On the Windows Taskbar,click on the KIC Host tab)
7.\x09Direct connection to SlimKIC2000,restart SlimKIC2000 Unit.
8.\x09Select Configure SlimKIC .
9.\x09Another Window will open.Select Write SlimKIC Setuop .You should see dialog appear in that window which says Write SlimKIC Successful.